LCV Specifications

 

Sample Point Capacity

10,750 data points when AC current and AC voltage are recorded together. 21,500 points when recording only AC current or only AC voltage.

Alarms

Over and under alarms for AC current and AC voltage.

Calibration

AC current and AC voltage inputs are factory calibrated. User single point offset calibration is available through software for both AC current and AC voltage.

Operating Temperature

32°F to 140°F (0°C to 60°C) using Alkaline Battery.

-40°F to 150°F (-40°C to 65°C) using Lithium Battery.

Time Accuracy

+/-100ppm @75°F

Relative Humidity Range

0% to 95%RH non condensing

Current Range

0 to 300 Amp AC

Current Accuracy

+/- 5%

Current Resolution

0.1 Amp AC

Voltage Range

0 to 500 Volts AC

Voltage Accuracy

+/- 1 VAC

Voltage Resolution

0.1 VAC

Dimensions

3’’ x 2.5" x 1.1’’ (7.6cm x 6.4cm x 2.8cm)

Weight

2.5 Oz (71g)

Power Source

9V Alkaline or Lithium battery.

Battery Life (average use)

18 Months using Alkaline Battery.

36 Months using Lithium Battery.