To set your LOGiT logger to start recording data, select Setup from the Logger menu. The setup options window is shown below.
· Sample Interval (Total Recording Length): Select the sample interval from a predefined list of intervals. The sample interval greatly affects the battery life of the logger.
· Custom Interval: Adds your own interval to the sample interval list.
· Enable Data Rollover: Check this box to have the logger continuously record the data, overwriting earliest recorded data. When this box is not checked, logging is stopped when the logger memory becomes full.
· Logging Control: Logging can start in three ways depending on the selection you make:
o Logging Starts When Door Is Closed, Stops When Door is Opened: Logging will not start until the sliding door is closed. Logging is stopped when the sliding door is opened. This mode does not allow real time monitoring of data. Logging must be stopped before the data can be retrieved. When the sliding door is opened the logger will flash the LED for 10 seconds, if the door is closed again during this interval, the logging will not stop.
o Logging Starts When Setup is Complete: Logging will start immediately once the Continue button is pressed, and OK is clicked on the follow-up warning message. When logging is started in this way, the sliding door does not have any effect on logging, but can still be used to secure the logger.
o Logging Starts At This Time: Logging will start at the user specified time. There is no limit how far in the future you can set the logging to start. When logging is started in this way, the sliding door does not have any effect on logging, but can still be used to secure the logger.
· Channels to Log: You can elect to disable unused logger channels. When a channel is disabled, its memory is allocated to the other channels, increasing the total logging time. The total logging time will be reflected in the Sample Interval window. Depending on the logger, some channels are required and cannot be disabled. At least one channel must be enabled for logging.
· User Information: Enter any text you like in this box. This field is used to identify the logger to the user. You can enter descriptive location, serial number, etc.
· Alarm Setup: Click this button to open the Alarm Setup Window.
· Advanced Setup: Click this button to open the Advanced Setup Window.