Sample Point Capacity |
10,750 points of thermocouple and ambient temperature. |
|
Alarms |
Over and under alarms for thermocouple and ambient temperature. |
|
Calibration |
Factory calibrated. User single point offset calibration is available through software for both thermocouple and ambient temperatures. |
|
Logger Operating Temperature |
32°F to 140°F (0°C to 60°C) |
|
Storage Temperature |
-40°F to 170°F (-40°C to 77°C) |
|
Real Time Clock Accuracy |
+/-100ppm @75°F |
|
Supported Thermocouples |
J, K, E, T, R, S Software selectable. |
|
Thermocouple Connector |
Industry standard miniature universal connector. |
|
Supplied Probe |
3ft FDA approved K type thermocouple probe. Probe temperature range is -40°F to 400°F (-40°C to 204°C) |
|
Thermocouple Sensor Accuracy |
Thermocouple wire dependant. The supplied probe accuracy is +/-4°F (2°C) or 0.75% of reading, whichever is greater. |
|
Thermocouple Sensor Resolution |
0.1°F (0.05°C) |
|
Internal Sensor Temperature Accuracy |
+/-2°F (1°C). For full accuracy allow a few minutes of thermal setting after connecting the thermocouple to the logger. |
|
Internal Sensor Temperature Resolution |
0.1°F (0.05°C) |
|
Dimensions |
3’’ x 2.5" x 1.1’’ (7.6cm x 6.4cm x 2.8cm) |
|
Weight |
2.5 Oz (71g) |
|
Power Source |
9V Alkaline or Lithium battery (Not Included). |
|
Battery Life (average use) |
20 Months using Alkaline Battery. 40 Months using Lithium Battery. |
|
Software Compatibility |
Compatible with LOGiTpc software version 1.20 and above. Software upgrades are available on the internet at www.supco.com |